Error Messages - Agilent Technologies 7250 Troubleshooting And Maintenance Manual

Accurate-mass quadrupole time-of-flight gc/ms system
Hide thumbs Also See for 7250:
Table of Contents

Advertisement

2
General Troubleshooting

Error Messages

Sometimes a problem in your MS will cause an error message to appear in the
MassHunter Workstation software. Some error messages appear only during
tuning. Other messages may appear during tuning or data acquisition.
Some error messages are "latched." These messages remain active in your data
system even if the condition that caused the message has corrected itself. If the
cause is removed, these messages can be removed by checking instrument
status through the data system.
Difficulty in mass filter electronics
Difficulty with the photo multiplier or
microchannel device
Difficulty with the fan
If a cooling fan fault occurs, the vacuum control electronics automatically shut
off the high-vacuum pump, the ion source, and mass filter heaters. Therefore, the
message: "The system is in vent state" may also appear. It is important to note
that even though the high-vacuum pump is off, the analyzer chamber may not
actually be vented. See
for precautions to take.
40
Pressure in the analyzer chamber is too high
RFPA is not adjusted correctly
Mass filter (quad) contacts are shorted or otherwise not working correctly
Mass filter is not working correctly
MS electronics are not working correctly
Broad peaks, such as the solvent peak, eluted while the analyzer was on
MS electronics are not working correctly
"The system is in vent state"
The fan is disconnected
The fan has failed
MS electronics are not working correctly
7250 Accurate-Mass Q-TOF Troubleshooting and Maintenance Manual
on page 43 in this section