6
Vacuum System

To replace the IDP pump HEPA filter cartridge

This procedure applies only to the dry scroll foreline pump.
Materials needed
The IDP pump operates at high temperatures. Do not touch any part until you
WAR N IN G
are sure it is cool.
The MS must be off and vented, and the foreline pump must be off, before
C AU T I O N
performing this procedure.
Procedure
1 Undo the three latches securing the top of the HEPA filter.
2 Remove the lid, and the filter cartridge from the filter.
3 Insert the new filter cartridge, and replace the lid on top of the filter.
4 Secure the lid using the three latches.

To change the exhaust silencer filter cartridge

Materials needed
7250 Accurate-Mass Q-TOF Troubleshooting and Maintenance Manual
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
HEPA filter with cartridge (SCRINTRPNW25)
HEPA filter cartridge (REPLHEPAFILTER1)
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Exhaust silencer filter cartridge
large (REPLSLRFILTER1)
107