To Position the GC Next to the MS

This procedure is used to position the GC next to the MS after moving the GC
away from the MS, or after relocating the instrument.
Ensure the GC/MS interface, GC inlet, and GC oven have cooled. These areas
WAR N IN G
can be hot enough to burn skin.
Procedure
1 Slide the units together and ensure you do not damage the capillary column.
2 Before closing the gap between the MS and the GC, connect the interface
3 Push the GC towards the MS, and close the gap.
4 If needed, connect the column to the transfer line. See the
5 Wind the excess column slack in the GC oven around the column basket.
6 Turn on the GC, and start carrier gas flow.
78
cable.
Accurate-Mass Q-TOF GC/MS System Operating
7250 Accurate-Mass Q-TOF Troubleshooting and Maintenance Manual
Agilent 7250
Manual.