6
Vacuum System
Exhaust silencer filter
KF Flange
KF Clamp wing nut
Figure 19.
To install the exhaust silencer filter
Procedure
This procedure applies to the rotary vane and dry scroll pumps. The dry scroll
pump is shown in the video.
1 Place the O-ring against the IDP-15 exhaust. (See
2 Place the exhaust silencer filter against the O-ring.
3 Place the clamp over the O-ring and KF flanges.
4 Tighten the clamp using the long screw and wing nut.
104
Foreline pump exhaust silencer filter
7250 Accurate-Mass Q-TOF Troubleshooting and Maintenance Manual
Figure 19
on page 104.)