Emerson Micro Motion 9739 MVD Configuration And Use Manual page 71

Hide thumbs Also See for Micro Motion 9739 MVD:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20016855
Option
Description
Informational Actions when fault is detected:
Actions when alert clears:
Ignore
No action
Status alerts and options for Status Alert Severity
Table 5-2: Status alerts and Status Alert Severity
Alert code
Status message
A001
EEPROM Error (Core
Processor)
A002
RAM Error (Core Processor) Fault
A003
No Sensor Response
A004
Temperature Overrange
A005
Mass Flow Rate Overrange
A006
Characterization Required
A008
Density Overrange
A009
Transmitter Initializing/
Warming Up
A010
Calibration Failure
A011
Zero Calibration Failed:
Low
A012
Zero Calibration Failed:
High
A013
Zero Calibration Failed:
Unstable
A014
Transmitter Failure
A016
Sensor RTD Failure
A020
No Flow Cal Value
A021
Incorrect Sensor Type (K1)
A026
Sensor/Transmitter
Communications Failure
A027
Security Breach
A102
Drive Overrange
Configuration and Use Manual
The alert is posted to the Alert List.
The status LED (if available) changes to red or yellow (depending on alert severity).
The status LED (if available) returns to green and may or may not flash.
Default severity Notes
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Informational
Configure device options and preferences
June 2019
Configurable?
No
No
Yes
No
Yes
Yes
Yes
Yes
No
Yes
Yes
Yes
No
Yes
Yes
No
No
No
Yes
71

Advertisement

Table of Contents
loading