Troubleshooting
June 2019
10.25.32 A111
Alert
Frequency Output Fixed
Cause
The Frequency Output has been configured to send a constant value.
Recommended actions
1. Cycling power to the meter or restarting totalizers will restore the Frequency Output to normal
operation.
2. Check whether the output is in loop test mode. If it is, unfix the output.
3. Check whether the output has been set to a constant value vial digital communication.
10.25.33 A112
Alert
Upgrade Transmitter Software
Cause
The transmitter software is down-level from the core processor software.
Recommended actions
Contact customer support.
10.25.34 A113
Alert
mA Output 2 Saturated
Cause
The calculated mA Output value is outside the configured range.
Recommended actions
1. Check the settings of Upper Range Value and Lower Range Value.
2. Check process conditions.
Actual conditions may be outside the normal conditions for which the output is configured.
3. Check for air in the flow tubes, tubes not filled, foreign material in the tubes, coating in the tubes, or
other process problems.
4. Verify that the measurement units are configured correctly for your application.
5. Purge the sensor tubes.