June 2019
Matrix name
Description
Deg Brix
Matrix represents a hydrometer scale
for sucrose solutions that indicates the
percent by mass of sucrose in solution
at a given temperature. For example,
40 kg of sucrose mixed with 60 kg of
water results in a 40 °Brix solution.
Deg Plato
Matrix represents percent extract, by
mass, in solution, based on °Plato. For
example, if a wort is 10 °Plato and the
extract in solution is 100% sucrose, the
extract is 10% of the total mass.
HFCS 42
Matrix represents a hydrometer scale
for HFCS 42 (high-fructose corn syrup)
solutions that indicates the percent by
mass of HFCS in solution.
HFCS 55
Matrix represents a hydrometer scale
for HFCS 55 (high-fructose corn syrup)
solutions that indicates the percent by
mass of HFCS in solution.
HFCS 90
Matrix represents a hydrometer scale
for HFCS 90 (high-fructose corn syrup)
solutions that indicates the percent by
mass of HFCS in solution.
4.8.4 Derived variables and calculated process variables
The concentration measurement application calculates a different set of process variables from each derived
variable. The process variables are then available for viewing or reporting.
Derived variable Description
Density
Mass/unit volume,
at Reference
corrected to a given
reference temperature
54
Density unit
3
g/cm
3
g/cm
3
g/cm
3
g/cm
3
g/cm
Calculated process variables
Density at
Standard
reference
volume
Specific
temp
flow rate
gravity
Configuration and Use Manual
MMI-20016855
Temperature
unit
Derived variable
°C
Mass
Concentration
(Density)
°F
Mass
Concentration
(Density)
°C
Mass
Concentration
(Density)
°C
Mass
Concentration
(Density)
°C
Mass
Concentration
(Density)
Concentra
Net mass
tion
flow rate
Micro Motion 9739 MVD Transmitters