Parameter Measurement - UNI-T UPO2102 User Manual

Digital phosphor oscilloscope
Hide thumbs Also See for UPO2102:
Table of Contents

Advertisement

User's Manual

8.1 Parameter Measurement

MSO/UPO2000 series oscilloscope can automatically measure 36 kinds of parameters. It includes voltage,
time and other.
① Voltage
Max:The voltage from the highest point of the waveform to GND.
Min:The voltage from the lowest point of the waveform to GND.
High:The voltage value from the flat top of the waveform to GND.
Low: The voltage value from the bottom of the waveform to GND.
Middle: Half of the sum of the voltage values at the top and bottom of the waveform
Pk-Pk: Peak-to-peak value, the voltage value from the highest point to the lowest point of the waveform.
Amp:Voltage from top to bottom of the waveform.
Mean:The average amplitude of the waveform in the screen.
CycMean:Average amplitude of the waveform in one period.
RMS: The effective value. The energy generated by the conversion of AC signal, it corresponds to the DC
voltage that generates equivalent energy.
CycRMS: The energy produced by the conversion of AC signal in a period, it corresponds to a DC voltage
that generates equivalent energy.
AC RMS: The RMS value is the waveform which DC component has removed.