UNI-T UPO2102 User Manual page 34

Digital phosphor oscilloscope
Hide thumbs Also See for UPO2102:
Table of Contents

Advertisement

User's Manual
MSO/UPO2000 Seri es
Figure 6-6
Based on sinθ=A/B or C/D, θ is the phase angle between channels, the definition of A, B, C, D see Figure 6-6.
So the phase angle isθ=±arcsin(A/B)or θ=±arcsin(C/D).
If the main spindle of elliptical within I, III quadrant, then the acquired phase angle should at I, IV quadrant,
that is within (0~π/2)or(3π/2~2π).
If the main spindle of elliptical within II, IV, then the acquired phase angle should within (π/2~π)or(π~
3π/2).
In addition, if the frequency or phase difference of the two signal to be measured are integer times,
calculating the frequency and phase relation of the two signals based on the Figure 6-7.
Figure 6-7
6.5 Multi-Scopes
In multi-scopes, CH1~CH4 can be set to different time base scale, so that signals of different frequency in
multiple channels can be observed at the same time. Enter Multi-Scopes interface via horizontal menu.
Switch Multi-scopes to enter this mode as shown in Figure 6-8. Each channel can be set to different
frequency, amplitude and waveform and all can be stable generated in multi-scopes.
In multi-scopes, it can turn on, off, activate channel. Time base scale, volts/div scale, horizontal
displacement, vertical displacement, trigger setting of channel can be adjusted independently.
Multi-Scopes supports split the screen to display the current channel on the upper part for observing
waveform, as shown in Figure 6-5.
34 / 137