Page 18
• Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
DHO4000 series digital oscilloscope is designed to meet the designing, debugging, and testing requirements of the mainstream oscilloscope market. Adopting the brand-new chipset "Centaurus" developed by RIGOL, this series achieves a fast waveform capture rate of 1,500,000 wfms/s with the UltraAcquire mode, 500 Mpts...
This manual gives you a quick overview of the front and rear panel, user interface as well as basic operation methods of DHO4000 series. http:// For the newest version of this manual, download it from RIGOL official website ( www.rigol.com Publication Number...
The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument. 2. Inspect the instrument In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative.
Setup > Language to select the system language. 4.3.6 Connecting Probes RIGOL provides passive and active probes for DHO4000 series. For specific probe DHO4000 series Data Sheet . For detailed technical information models, please refer to of the probes, please refer to the corresponding Probe User Guide.
Page 41
USB-GPIB module. - USB storage device: imports or exports data (software update, waveform, setup, or captured image). - USB-GPIB module: extends the GPIB interface for RIGOL instruments that integrates the USB HOST interface but not the GPIB interface. 15. Power Switch Press the switch to power on/off the oscilloscope.
(used for obtaining desired the option license code). Then, you can install the option according to the following steps. 1. Obtain an option license http://www.rigol.com a. Log in to the RIGOL official website ( ), click SERVICE CENTRE > License Activation to enter the "Registered product license code" interface.
20 MHz • For DHO4404 and DHO4804, when the vertical scale is less than or equal to 500 μV, you can only set the bandwidth limit to 20 MHz or 250 MHz, but cannot turn off the limit; when the vertical scale is less than or equal to 200 μV, the bandwidth limit is fixed to 20 MHz.
In the trigger debugging process of video signals, the frequency in different part of the signal can be reflected by a different brightness, as RIGOL's digital oscilloscope provides the intensity graded color display function. Experienced users can quickly judge the signal quality and discover abnormalities during the debugging process.
1. File Header There is only one file header in a binary file. The file header contains the following information. Table 20.2 File Header Two-byte characters, RG, indicating that the file is the RIGOL Cookie binary data file format. Version Two-byte, indicating the file version.
SCPI (Standard Commands for Programmable Instruments) commands. This chapter will illustrate how to use the RIGOL Ultra Sigma software to remotely control the instrument via various interfaces. Note: When communicating with the PC via GPIB, the instrument does not support large data transmission operation such as screenshot and waveform reading.
PC via the USB interface. You can also click USB-TMC to search for the resource. 3. View the device resource The resources found will appear under the "RIGOL Online Resource" directory, and the model number and USB interface information of the instrument will also be displayed.
4. View the device resource The resources found will appear under the "RIGOL Online Resource" directory. 5. Control the instrument remotely Right-click the device resource name and select "SCPI Panel Control" to open the remotely command control panel.
5. View the device resource Click OK to go back to the main interface of Ultra Sigma. The searched instrument resource will be displayed under the directory of "RIGOL Online Resource". 6. Control the instrument remotely Right-click the device resource name. In the displayed menu, select "SCPI Panel Control"...
After restarting the instrument, insert the USB storage device again to check whether it can work normally. d. If the USB storage device still cannot work normally, please contact RIGOL. 4. The touch-enabled operation does not work. a. Check whether you have locked the touch screen. If yes, unlock the touch screen.
RIGOL guarantees free replacement or repair for the defective product. To get repair service, please contact your nearest RIGOL sales or service office. There is no other warranty, expressed or implied, except such as is expressly set forth herein or other applicable warranty card.