4 Product Storage and Handling Requirements

4.1 Product Storage and Handling limits

To obtain Temperature, Humidity or Pressure requirements for components not specified in
Table
1-4, refer to the appropriate component Pre–Installation Manual listed in
Innova LC System
Table 1-4: Storage and Transport
Component
All UPS units (see also
tion
4.2)
All Monitors
All Injectors
All Footswitches
Vivid E9 Option
Cardiac Collimator
Diamentor
All other components
NOTE:
(1) The detector should be stored at 10 to 40 °C (50 to 104 °F) and less than or equal
to 90% RH in the plastic wrapped shipping box. (This should include two bags of
desiccant as well). The lowest temperature (e.g. 10 ºC (50 °F)) and humidity is
preferable. If they are to be stored outside of their shipping box or in the inner
shipping box without plastic wrapping they should be stored at 20 °C (68 °F) or less
and 30% RH or less. In terms of transportation, do not expose to temperatures below
–20 ºC (–4 °F) in its shipping box for more than 15 hours. The detector will reach the
ambient temperature after 20 to 25 hours. The detector should not be allowed to
reach temperatures less than –10 ºC (14 °F) or irreparable damage to the detectors
scintillator will occur. Care must be taken when removing a detector from a shipping
box. If the detector has been subject to cold temperatures for an extended period the
detector in the box should be allowed to sit in the plastic wrapped box to reach room
temperature. This will prevent condensation from occurring. Condensation on the
detector can cause irreparable damage to the electronics. Storage 10 to 40 ºC (50 to
104 °F); 10 to 90 % RH, 250 day storage transportation –20 to +60 ºC (–4 to 140 °F)
and 10 to 80% RH. The Detector chiller is shipped within GEMS packaging.
Chapter 1 General Requirements
Innova
IGS 520, Innova
TM
NOTICE
Avoid extremes in temperatures
Compatibility.
TEMPERATURE
MIN
Sec‐
-20°C (-4°F)
-20°C (-4°F)
-20°C (-4°F)
-20°C (-4°F)
-40°C (-40°F)
-20°C (-4°F)
-20°C (-4°F)
-40°C (-40°F)
IGS 530, Innova
TM
HUMIDITY (see Note (2))
MAX
MIN
+40°C (+104°F)
10%
+55°C (+131°F)
10%
+60°C (+140°F)
10%
+60°C (+140°F)
10%
+70°C (+158°F)
30%
+70°C (+158°F)
10%
+60°C (+140°F)
0%
+70°C (+158°F)
10%
IGS 540 Pre-Installation Manual
TM
Direction 5499972-1-1EN, Revision 2
Chapter 2, Basic
PRESSURE
MAX
MIN
90%
79.4 kPa
80%
79.4 kPa
80%
79.4 kPa
90%
79.4 kPa
60%
79.4 kPa
95%
79.4 kPa
85%
79.4 kPa
95%
79.4 kPa
MAX
106 kPa
106 kPa
106 kPa
106 kPa
106 kPa
106 kPa
106 kPa
106 kPa
43