Software Error - Air Techniques ScanX Swift View Installation And Operating Instructions Manual

Hide thumbs Also See for ScanX Swift View:
Table of Contents

Advertisement

Available languages

Available languages

Troubleshooting
EN-
Error
US
Lamination of the PSP
detaches at the edge
14.2

Software error

Error
"Too much ambient light"
"Incorrect power supply unit"
"Overtemperature"
"Erasure unit fault"
Imaging software fails to rec-
ognize the unit
44
Possible cause
Wrong retainer system used
PSP handled incorrectly.
Possible cause
Unit is exposed to too much
light
Incorrect power supply unit con-
nected
Laser or erasure unit too hot
LED defective
Unit not switched on
Connecting cable between unit
and computer not correctly con-
nected
Computer does not detect any
connection to the unit
Hardware error
The IP address of the unit is
being used by another unit
Remedy
Only use original PSP and film
retainer systems.
Use the PSP correctly.
Comply with the operating
instructions of the PSP and
film retainer systems.
Remedy
Darken the room.
Turn the unit such that no light
can directly enter into the
entry slot.
Use the enclosed power sup-
ply unit.
Switch the unit off and allow it
to cool.
Contact technician.
Switch the device on.
Check the connecting cable.
Check the connecting cable.
Check the network settings
(IP address and subnet
mask).
Contact technician.
Check the network settings
(IP address and subnet mask)
and assign a unique IP
address to each unit.
Inform a service technician, if
the problem persists.
2142100008L41 2202V007

Advertisement

Table of Contents
loading

This manual is also suitable for:

G8500a

Table of Contents