Installation
EN-
The network connection is used to exchange
information or control signals between the unit
US
and a software installed on a computer, in order
to, e. g.:
– Display parameters
– Select operating modes
– Indicate messages and error situations
– Change device settings
– Activate test functions
– Transmit data for archiving
– Provide documents concerning the devices
The unit can be connected to the network with a
network cable or via WLAN.
For information on connection via WLAN,
see "Configuring WLAN on the unit".
Combining devices safely
– Safety and essential performance features are
independent of the network. The device is
designed appropriately for operation inde-
pendent of a network. However, some of the
functions are not available in this case.
– Faulty manual configuration can lead to signifi-
cant network problems. The expert knowledge
of a network administrator is required for con-
figuration.
– The data connection utilizes part of the band-
width of the network. Interactions with other
medical devices cannot be completely
excluded. Apply the IEC 80001‑1 standard for
risk assessment.
– The device is not suitable to be connected
directly to the public internet.
Take care when connecting units together or to
parts of other systems as there is always an ele-
ment of risk (e.g. due to leakage currents).
Only connect units when there can be no
question of danger to operator or to patient.
Only connect units when it is safe to do so and
when there is no risk of damage or harm to the
surroundings.
If it is not completely clear from the data sheet
of the unit that such connections can be safely
made or if you are in any doubt, always get a
suitably qualified person (e.g. the relevant man-
ufacturer) to verify that the setup is safe.
Only connect peripheral units (e.g., computer,
monitor, printer) that conform at least to the
requirements set out in IEC 60950‑1.
20
Connecting the unit via the network cable
Remove the cover from the rear of the device.
Plug the enclosed network cable into the net-
work socket of the device.
Replace the cover.
The cover on the rear must be correctly
fitted when the device is operated within
the patient environment.