EN-
10 Correct use of phosphor
US
storage plates
WARNING
Risk of cross contamination when not
using the barrier envelope or when
using the barrier envelope more than
once
Do not use an phosphor storage plate
without a barrier envelope.
Do not re-use the barrier envelope
(disposable item).
CAUTION
Image data on the phosphor storage
plate is not permanent
The image data is altered by light, natu-
ral X-ray radiation and scattered X-ray
radiation. This will lead to a reduction in
diagnostic information and clarity.
Read the image data within 30 minutes
of exposure.
Never handle exposed phosphor stor-
age plates without the barrier enve-
lope.
Do not subject an exposed phosphor
storage plate to X-ray radiation before
and during the scanning process. Do
not X-ray during the scanning process
if the unit is in the same room as the X-
ray tube.
Phosphor storage plates must only be
read using a phosphor storage plate
scanner that is approved by Air Techni-
ques.
28
CAUTION
Phosphor storage plates (PSP) are
toxic
PSPs that are not used with a barrier
envelope can lead to poisoning when
placed in the mouth or swallowed.
Only place PSPs in the patient's mouth
in a barrier envelope.
Do not swallow the PSP or parts of it.
If the PSP or parts of it have been
swallowed, consult a specialist doctor
immediately and remove the PSP.
If the barrier envelope was damaged in
the patient's mouth, rinse the mouth
thoroughly with copious amounts of
water. Do not swallow the water in the
process.
Phosphor storage plates are flexible like X-ray
film. However, the phosphor storage plates
should not be kinked.
Do not scratch the phosphor storage plates.
Do not subject the phosphor storage plates to
pressure from hard or pointed objects.
Do not soil the phosphor storage plates.