Barrier Envelope; Phosphor Storage Plate - Air Techniques ScanX Swift View Installation And Operating Instructions Manual

Hide thumbs Also See for ScanX Swift View:
Table of Contents

Advertisement

Available languages

Available languages

Usage
EN-
Input unit
US
The input unit must be cleaned and disinfected if
there are indications of contamination or visible
soiling.
NOTICE
Heat can damage plastic parts
Do not treat any part of the unit in a
thermal disinfector or steam sterilizer.
Tap
on the touch screen.
The fixing mechanism moves into the cleaning
position.
Press the release key and remove the cover
upwards at the same time.
1
Remove the fixing mechanism by moving it
upwards.
36
2
Clean the cover, fixing mechanism and inside
parts with a soft, lint-free cloth that has been
dampened with cold tap water.
Disinfect the cover, fixing mechanism and
inside parts with a disinfection wipe.
Alternatively, use a spray disinfectant on a soft,
lint-free cloth. Comply with the operating
instructions of the disinfectant.
The cover and fixing mechanism can also be
disinfected in an immersion disinfection sys-
tem.
Remount the fixing mechanism.
Remount the cover.
Touch OK on the touch screen.
The fixing mechanism moves into the starting
position.
12.2

Barrier envelope

Clean the Barrier Envelope after being removed
from the patient's mouth with a disinfectant
wipe, such as Air Techniques Monarch Surface
Wipes.
Allow the Barrier Envelope to completely dry
prior to ejecting the PSP.
12.3

Phosphor storage plate

Cleaning and disinfectant wipes are unsuitable
for the cleaning of phosphor storage plates and
may damage them.
Only use a cleaning agent that is compatible with
the materials:
Air Techniques recommends the PSP Cleaning
Wipes (see "3.4 Consumables"). Only this prod-
2142100008L41 2202V007