Error
Error code 1026
Error code 1100
Error code 1104
Error code -1116
Error code -1117
Error code -1118
Error code -1121
Error code 1153
Error code 1154
Error code 1160
Error code -1171
Error code 1172
Error code 10000
Error code 10009
2142100008L41 2202V007
Possible cause
Incorrect acquisition mode
Permitted time for scan process
exceeded
Erasure unit fault
Drive feed blocked
Feed position error
Input unit cover open
Input unit fixing mechanism
missing
Unit fault
Internal data communication
fault
Final radiation deflector rotation
speed not reached
Fault on laser
SOL sensor timeout
Fault on the laser, SOL sensor or
pentaprism assembly
Unit is exposed to too much
light
Internal communication error
warning; unit remains ready for
operation
Troubleshooting
Remedy
Select a different acquisition
mode
Inform a service technician.
Update the firmware.
Reset the scanning modes to
the factory settings via the
unit interface or the Imaging
Software.
Contact technician.
Check the belt drive.
Check for blockage, remove
PSP from unit.
Contact technician.
Replace the erasure unit.
Remove the blockage.
Contact technician.
Contact technician.
Check the feed (ease of
movement, drive belts).
Close the cover.
Deactivate cleaning mode.
Insert the fixing mechanism.
Deactivate cleaning mode.
Switch the unit off and on
again.
Update firmware.
Switch the unit off and on
again.
Update firmware.
Contact technician.
Update firmware.
Replace the radiation deflec-
tor subassembly if the prob-
lem occurs regularly.
Return the unit for repair.
Inform a service technician.
Update the firmware.
Darken the room.
Turn the unit such that no light
can be directly incident in the
entry slot.
Update the firmware.
EN-