. Furthermore, only the accessories described in the installation and operating instructions or approved by Air Techniques for use with this unit may be used . If other accessories are used, Air Techniques can assume no warranty for the safe operation and reliable function of the unit .
Important Information 1.5 Incorrect Usage. Any usage of the ScanX Swift that is not described here as correct usage is considered as incorrect usage . The manufacturer is not liable for any damages arising from incorrect usage . The operator bears all risks and liabilities .
Wear protective gloves 3.2 ScanX Swift. The model identification plate can be found on the rear of the ScanX Swift . The following terminology and/ or symbols are used in conjunction with this appliance, model identification plate and power unit: Warning dangerous voltage.
. Save all cartons and packing materials to protect ScanX Swift in the event that it is to be transported or shipped in the future . ScanX Swift consists of the main assembly and accessory kit as listed above . Verify that all listed items were received . If any item is missing, notify your dealer .
These wireless flexible sensors are durable and reusable thousands of times when properly handled . Upon exposure to X-rays, the plate stores a latent image, which is scanned by the ScanX Swift . After scanning, the image is processed via the user-supplied computer running authorized software and ready for viewing in seconds .
Operating power control switch On / Off button Switch on: Press the On / Off button. Press the On / Off button for 2 seconds. ScanX Swift shuts down, the Switch off: green and blue LEDs blink alternately. Status LED indicators Displays information on status of plate transport.
Important Information Computer System Requirements ScanX Swift must be connected to a compliant Computer System supplied by the customer. Optional Imaging Software (PACS) can be used for enhanced features. Smooth operation of Air Techniques hardware and software can be affected through specific hardware and software incompatibility on site at the client despite meeting system requirements.
8.2 Counter Top/Table Set-up. ScanX Swift may be located almost anywhere in the office . Follow these guidelines for optimum performance: Provide a stable, flat counter top large enough to hold the scanner .
ScanX Swift. 9. Software Installation Before connecting ScanX Swift to your computer or attempting to use it for the first time, run the Setup program on the Drivers and Utilities Disk . This CD contains the software necessary to integrate ScanX Swift with your imaging software .
Page 13
Unit can not be operated using USB 1.1. Swift can not operate using USB 1 .1 . Use USB 2 .0 or higher . Figure 10-1 . ScanX Swift Rear Panel Connection Ports IMPORTANT: Make sure Adapter Connector is properly installed...
• Only use Air Techniques Plates. ScanX Swift is operated in combination with Air Techniques plates in size #0, #1 and #2 . (See 4 .1, System Components .) Each size of plate requires the matching barrier envelope . Exposure of the...
Printed Side Inactive Side Active Sensitive Side Figure 12-1 . Typical Air Techniques Size #2 Plate Configuration 12.2 Working with Plates. Plates must be visibly checked for signs of damage . Damaged plates (e .g . a torn outer layer, or visible scratches, which will affect the diagnostic quality) must be replaced .
. Peel off the adhesive strip and seal the envelope . See Figure 12-4 . View A . View B . View C . Insert the Plate . Peel off the adhesive strip . Seal the Barrier Envelope . Figure 12-4 . Sealing the Plate in the Barrier Envelope Page 16 Air Techniques, Inc.
13. Plate Scanning with Swift 13.1 Changing the Inlet Guide. ScanX Swift can be used to scan plates Size #0, Size #1 and Size #2 . Each size of plate requires the matching size inlet guide . Size #0 guide accepts Size #0 PSPs ...
13.3 Scanner Plate Processing Options. Using the ScanX Swift, the user has the option to perform single plate processing or batch processing depending on their needs . The difference between the two processing procedures is mainly how the plates are handled . Refer to paragraphs 13 .4 and 13 .5 to perform single plate processing . Paragraphs 13 .6 and 13 .7 provide the batch processing procedures .
Use a new Barrier Film for each patient. 13.4 Single Plate Processing Scanner Preparation Apply a barrier film to the ScanX Swift input guide making sure to align the opening in the film center with the slot of the inlet guide .
OVER IMMEDIATELY. Failure to do so may result in erasure of the PSP . 3 . Lay the box on a flat surface near the ScanX Swift as shown by Figure 13-4, View A . 4 . As shown by Figure 13-4, View B, remove the exposed PSP from its washed and dried envelope as follows: a .
The data on the plate has not been (completely) erased due to a malfunction of the ScanX Swift . The special Erase mode enables the erasure unit in ScanX Swift to be activated alone. No image data will be transmitted to the View C software.
Do not allow fluids to enter the unit. Where the surfaces of the appliance show signs of contamination or visible dirt, clean and disinfect the unit . Switch ScanX Swift off by pressing the ON/OFF switch for 2 seconds .
• Inform your Service Technician eration after switching on 30 seconds, then the motor is defective 5. Image editing software • ScanX Swift not switched on • Switch on ScanX Swift does not recognise • Connecting cable between • Check the USB cable...
Page 25
14 Pale stripes in image • Too much stray light entered • Darken the room during scanning • Rotate the ScanX Swift so that no direct light falls on the film entry slot 15. Blue LED illuminated • No connection between •...
Never power down the system during a scanning session. 18.1 Powering Down System. ScanX Swift is designed to be left on continuously during the active day . At the end of the day, or whenever desired, power down the system as follows: 1 .
Viewing - provides automatic image mounting that can be customize to the user’s needs . Flexibility - includes a built-in bridging solution to over 35 Practice Management Software titles . VISIX is the superb complement for ScanX Swift . Contact your authorized dealer for further information . Air Techniques, Inc. Page 27...
The file sizes here are approximate sizes which have been rounded up . The file sizes can be considerably reduced by using a suitable compression program . Scan mode Intraoral Very High Intraoral Standard Theor. Resolution LP/mm 23.8 Pixel size µm Size #0 Size #2 Page 28 Air Techniques, Inc.
ScanX Swift is warranted to be free from defects in material and workmanship from the date of installation for a period of 2 years (24 months) . ScanX Swift is designed solely for use in a dental office environment and this warranty is not applicable to other applications .
Page 32
For over 50 years, Air Techniques has been a leading innovator and manufacturer of dental products . Our priority is ensuring complete satisfaction by manufacturing reliable products and providing excellent customer and technical support . Whether the need is digital imaging, utility room equipment or merchandise, Air Techniques can provide the solution via our network of authorized professional dealers .