If a scanning job is started via the touch screen,
the image is saved to the memory card and
US
needs to be transferred to the computer later on.
After scanning, the phosphor storage plate runs
through the erasure unit. Image data still held on
the phosphor storage plate is erased with the aid
of bright light.
The phosphor storage plate is then ejected for
re-use.
Operating elements
1
Touch screen
2
On/off switch
The touch screen allows the unit to be operated
when it is not connected to a computer. Instruc-
tions can be entered on the touch screen either
with the tip of a finger or the stylus.
The Help button can be used to open a help
page for the respective page. The Messages
button can be used to call up current messages.
Connections
The connections are located on the rear of the
unit, underneath the cover.
1
Connection for power supply unit
2
Reset button
3
USB connection (for WLAN stick)
4
Network connection status LEDs
5
Network connection
6
AUX connection for diagnostic units
14
ScanManager
When the ScanManager is enabled, more than
one X-ray job can be transmitted simultaneously
to the unit from different computers. The unit
manages the X-ray jobs in a queue from which
the respective X-ray job can be selected using
the touch screen and then executed.
Without ScanManager, the unit processes one X-
ray job at a time and is blocked until this job has
been completed. During this time, no further X-
ray jobs can be transmitted to the unit from other
1
computers.
ScanManager can be enabled via Set-
tings > System Settings> Operating
2
Type.
5.2
Phosphor storage plate
The PSP stores X-ray energy, which is re-emitted
in the form of light after excitation by the laser.
This light is then converted into image informa-
tion in the phosphor storage plate scanner.
The PSP has an active side and an inactive side.
The PSP must always be exposed on the active
side.
When used properly, a PSP can be exposed,
read and erased several hundred times provided
there is no mechanical damage. The PSP must
be replaced if there are any signs of damage, e.g.
if the protective layer is damaged or there are
visible scratches that could interfere with the
diagnostics.
1
2
3
4
5
6
2142100008L41 2202V007