Section 9
Maintenance Schedule
Figure 14: Maintenance kit assembly
detail
68
Thermo Scientific Orion
PRIME FLUIDICS menu the (-) sign for the priming volume will be reset to (+)
NOTE:
upon next menu entry.
SAMPLE/STANDARD
PUMP (larger tubing OD)
Always use 2 drops of Silicone Oil on the platen before replacing on pump
Reagents 1, 2 and 3 Pumps
(smaller tubing OD, all the same)
2230XP Silica Analyzer