Section 9
Maintenance Schedule
74
Thermo Scientific Orion
Measurement/Calibration/Validation Messages
Messages
Testing optics, low signal
level . . .
Cell may be coated and may
require cleaning.
Testing optics please check
for stray light
Please check for heater
failure
Checking the reaction
temperature. Temperature
below set point.
Checking the reaction
temperature. Temperature
above set point.
Taking blank measurement,
unstable signal.
Collecting data, unstable
signal
Messages
Unstable signal, stray light is
high please check.
Optimum light level.
Calibration is aborted.
Messages
Temperature below set point
please check for hardware
malfunction.
Temperature above set point
please check for hardware
malfunction.
Cause
Low light output from Optics.
Optical window of cell may be coated.
External light is entering the cell.
If heater is enabled in Method, and the
Set point temperature is higher than
ambient and Cell temperature differs
from set point by ± 10ºC for 90 secs..
If the temperature is not able to reach
the set point and is below the set
point.
If the temperature is not able to reach
the set point and is above the set
point.
Unstable temperature, Cell may be
coated, or may be due to Optics
failure.
Unstable temperature or cell may
be coated, or may be due to optics
failure.
Calibrating Optics Messages
Cause
Some external Light is entering in to
the cell.
Cell may be coated.
Calibrating Thermistor Messages
Cause
May be due to the connectors or the
thermistor or heater malfunctioning.
May be due to the connectors or the
thermistor or heater malfunctioning.
Solution
Run a Clean Cycle and try to
clean the reaction cell with
detergent.
Check if overflow cap on top of
the reaction cell is loose.
Try a different set point
temperature. Call a service agent.
Re-calibrate the Thermistor.
Try a different set point
temperature. Call a service agent.
Re-calibrate the Thermistor.
Try a different set point
temperature. Call a service agent.
Perform Clean Cycle and
Calibrate Optics and thermistor.
Perform Clean Cycle and
calibrate optics and thermistor.
Solution
Check for possible external light
entering areas on the reaction
cell.
Perform Clean Cycle
and then Calibrate Optics.
Solution
Check for thermistor and heater
connections, and calibrate them.
Check for thermistor and heater
connections, and calibrate them.
2230XP Silica Analyzer