Thermo Scientific Orion 2230XP User Manual page 26

Silica analyzer
Table of Contents

Advertisement

Section 3
Installation Environmental Considerations
Table 2: Main terminal strip pin
assignments
26
Thermo Scientific Orion
Terminal
Designation
Number
1
in1
2
GnD
3
in2
4
GnD
5
in3
6
GnD
7
in4
8
GnD
9
in5
10
GnD
11
in6
12
rTn
13
in7
14
rTn
15
+24V
16
GnD
17
Earth Ground
18
GnD
19
OUT1
20
Earth Ground
21
GnD
22
OUT2
23
nC1
24
nO1
25
C1
26
nC2
27
nO2
28
C2
29
nC3
30
nO3
31
C3
32
nC4
33
nO4
34
C4
Function
Digital input 1 Live (Level sensor 1 live)
Digital input 1 ground (Level sensor 1 ground)
Digital input 2 live (Level sensor 2 live)
Digital input 2 ground (Level sensor 2 ground)
Digital input 3 live (Level sensor 3 live)
Digital input 3 ground (Level sensor 3 ground)
Digital input 4 live (Level sensor 4 live)
Digital input 4 ground (Level sensor 4 ground)
Digital input 5 live (Level sensor 5 live)
Digital input 5 ground (Level sensor 5 ground)
Digital input 6 live (External 1 control)
Digital input 6 return (External 1 control return)
Digital input 7 live (External 2 control)
Digital input 7 return (External 2 control return)
24V DC live (limited to 20 mA source)
24V DC ground
System earth ground
Analog output 1 ground
Analog output 1 (900 Ohms Max)
System earth ground
Analog output 2 ground
Analog output 2 (900 Ohms max)
Normally closed contact relay 1
Normally open contact relay 1
Common contact relay 1
Normally closed contact relay 2
Normally open contact relay 2
Common contact relay 2
Normally closed contact relay 3
Normally open contact relay 3
Common contact relay 3
Normally closed contact relay 4
Normally open contact relay 4
Common contact relay 4