2230XP Silica Analyzer
Terms
Help message for the user
Display Help
to navigate between the menus.
Message showing the status
Display Info
of the current cycle performing.
Selectable precision to show the
Results Decimal Point
analysis result.
To display the result with large
Large Fonts Screen
fonts.
CALIBRATION PARAMETERS
The concentration of the standard
Standard
used for calibration.
Mode represents the calibration
Mode
type.
The time duration between two
Frequency
calibration cycles.
Calibrations start time, except for
Start Time
manual and editable frequency.
The CURRENT Slope value
represents the coefficient that is
Slope
multiplied by the absorbance to
calculate the concentration of the
sample +/- Intercept value.
Reference and current numbers
Intercept
based on the base line calibration
cycle.
Repeat tolerance tested during auto
Calibration Tolerance
calibration cycles to accept or reject
automatic calibrations.
VALIDATION PARAMETERS
Validation Source
Fluid source for validation.
Number of times that the validation
Number of Repeats
is to be performed.
ANALYSIS PARAMETERS
Mode
Instrument RESTART mode.
In the RUN mode, the frequency
Measuring Frequency
is the time duration between two
analysis cycles.
Measuring Units
Units for the analyzed sample.
Analysis Label
The rate at which the sample
Cell Wash Flow Rate
is pumped to wash the cell.
Description
Enabled, Disabled
Enabled, Disabled
XXXX.X, XXX.XX, XX.XXX, X.XXX,
XXXXX, Auto
Enabled, Disabled
0 to XXXXX (XXXXX Value depends on
precision selected in Setup ➔ Display
parameters ➔ result Decimal Point
Single-Point, Base-Line
Every 6Hr, 12Hr, 24Hr, Sun, Mon, Tue,
wed, Thu, Fri, Sat, Manual and Editable
0Hr to 99.99Hr
Depends on Time format in Setup ➔
Time Menu
Default value of -1
Default value of +/- 5%
Sample, Standard, Grab Sample, Not
Assigned
0 to 8
Auto Restart, Manual Restart
15, 30, 45 minutes, 1 hr, 0.20 hr are
selectable and editable minutes range
00.00 to 99.99Hr
Six characters long. Alphabets,
numerals and special symbols are
allowed.
Six characters long. Alphabets,
Numerals and +, - are allowed.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Terms Definition
Section 8
Possible Values, Options