2230XP Silica Analyzer
Terms
Prints the date, time and
Online Print
concentration after each analysis
cycle.
EDIT CALIBRATED VALUES
To change the flow rate of the
Pump1, Pump2, Pump3
pumps after their calibration.
The level detected as dark light
Optics Dark
level.
The level used as reference
Optics Xref
potential for light energy level.
PRIME FLUIDICS
The volume of the fluid selected for
Priming Volume
priming.
PUMPS, VALVES & STIRRER
Pump1, Pump2, Pump3
To check the functionality of pumps.
Valve1, Valve2, Valve3
To check the functionality of valves.
Stirrer
To check the functionality of stirrer.
RELAYS
Relay1, Relay2, Relay3,
To check the functionality of relays.
Relay4
ANALOG OUTPUTS
To test the functionality of analog
Channel1, Channel2
outputs for different selectable
currents.
LED & ADC
Select LED
To test the functionality of LED.
Intensity
Intensity of the LED.
Optics Amplifier Gain
Changing the gain of the amplifier.
Description
Enabled, Disabled
0.10 to 60.00 mL
0 to 300 mV
0 to 2400.00 mV
0 to 99.9 mL
ON, OFF
ON, OFF
ON, OFF
ON, OFF
0 mA, 4 mA, 10 mA, 12 mA, 20 mA
NIRx 810, None
0% to 100%
0 to 255
Terms Definition
Section 8
Possible Values, Options
Thermo Scientific Orion
65