Section 8
Terms Definition
62
Thermo Scientific Orion
Terms
The rate at which the reagent
Reagent 1–3 Flow Rate
is pumped to wash the cell.
Defines the stability level of the data
before accepting it. The stability
criteria is a measure of stability of
Stability Criteria
the absorbance level rate of 0.001
AU per sec as specified by the
number 1 thru 10.
Washing the cell after completion of
Post Wash Event
the event selected with the following
source, volume and stir time.
Fluid for post washing the reaction
Post Wash Cycle Source
cell.
Post Wash Cycle
The amount of fluid pumped
Volume
for post washing.
Post Wash Cycle Stir
The amount of time the fluid is
Time
stirred for post washing the cell.
CLEANER PARAMETERS
Cleaner Source
Fluid for cleaning the cell.
Volume to add
The amount pumped for cleaning.
The rate at which the cleaner
Clean Flow Rate
source is pumped to clean the cell.
The amount of time the fluid is
Stir Time
stirred for cleaning the cell.
The time duration between two
Frequency
cleaning cycles.
For weekly cleaning the day of
Start Day/Date
cleaning. For monthly cleaning the
date of cleaning.
The time to start the cleaning on the
Start Time
day or date selected.
TEMPERATURE UNITS
Units
Units of temperature.
METHOD PARAMETERS
The amount of sample pumped to
Cell Wash Volume
wash the cell.
Reaction Temperature
PID temperature control.
Control
Reaction Temperature
The temperature set point.
Wavelength
The LED wavelength.
Allows the blank measurement
Blank Measurement
in the analysis cycle.
The amount of reagent pumped
Reagent Volume
in to the cell.
Description
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
1 thru 10.
Analysis, Calibration, Validation, Not
Assigned.
Sample, Reagent, Cleaner, Standard
-99.9 to +99.9 mL
0 to 999 seconds
Sample, Reagent, Cleaner, Standard,
Not Assigned
0 to 255 mL
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
0 to 999 seconds
Manual, Daily, Weekly, Monthly
Day - Week days
Date - Current month dates (1 to 31)
00:00 hr to 23:59 hr
Celsius, Fahrenheit
1 to 100 mL
Enable, Disable
0 to 60
NIR3 810 nm
Disable, Before Reagent1, Reagent2,
Before Reagent3
0 to 6.00 mL
Possible Values, Options
2230XP Silica Analyzer