Section 9
Maintenance Schedule
Table 3: Error messages
72
Thermo Scientific Orion
The system will switch to Stand-By after 60 minutes of no keypad activity. In Stand-By, the
analyzer will rotate the pumps and exercise the pinch valves periodically to avoid pinched
tubing from sealing.
Extended Shutdown
To avoid clogging, flush the entire system with distilled water before shutdown. Disconnect and
empty out the reagents bottles and flush the lines using FLUIDICS PRIME menu. Release the
pressure on the peristaltic pump tubing and remove the tubes from the pinch valves.
Use the table below to diagnose and correct analyzer problems.
Messages
Login as Technician or
Operator
Password mismatch, Enter to
continue
Measurement is in progress
User programmable
measuring frequency
User programmable
calibration frequency
Two or more inputs are of
same type
Error in date, press F1 to
continue
Error in hours, press F1 to
continue
More than 6 users are not
allowed, delete a user to add
further
No records to delete.
Are you sure you want to
delete all analysis results?
Are you sure you want to
delete all calibration results?
Cause
User has not logged in, and trying to
access restricted operations.
User is logged in, but does not have
adequate privileges to access the
restricted operations.
Incorrect password has been entered.
Trying to perform measurement
when it is already under way.
Editing option of analysis frequency is
selected.
If same type of solution is assigned to
two or more inputs.
If same type of solution is assigned to
two or more input.
Date entered is invalid.
Hours entered is invalid.
Technician trying to add more than 6
users.
User is trying to delete
Calibration/Analysis records, when
there are no records stored.
All analysis results? User is trying to
delete all analysis records. (At least
one analysis record is available in the
database.)
User is trying to delete all calibration
records. (At least one calibration
record is available in the database.)
Solution
User has to login either as a
Technician or Operator to access
the restricted operations.
User is not authorized
to access the restricted
operations.
Press "Enter" to cancel the wrong
entry and enable re-entry of
correct password. If password is
lost consult factory for backdoor
access code.
Wait for measurement to
complete.
Edit the analysis frequency.
Edit the calibration frequency.
Reassign solution to inputs.
Press F1 and re-enter valid date.
Press F1 and re-enter valid hours.
More than 6 users are not
permitted at any time. To add a
new user after 6, delete an
existing user.
None.
None. Deleted records are not
recoverable.
None. Deleted records are not
recoverable.
2230XP Silica Analyzer