Setup Menu - Thermo Scientific Orion 2230XP User Manual

Silica analyzer
Table of Contents

Advertisement

Section 7
Menu Overview
50
Thermo Scientific Orion

Setup Menu

Alarms Menu
Allows configuration of alarm parameters (type, value and hysteresis). Also allows assignment
of individual alarms to faults.
Relays Mapping Menu
Allows mapping of multiple alarms to individual relays.
Analog Outputs Menu
Allows configuration of individual analog outputs.
Inputs Menu
Allows configuration of five internal inputs (for fluidics sensors) and two external inputs (for
instrument control).
Display Parameters Menu
Allows configuration of display-related parameters.
Calibration Parameters Menu
Allows configuration of parameters for calibration. The menu shows two columns tagged as
REFERENCE and CURRENT. The parameters listed under the REFERENCE column are set
for validating auto calibrations repeats. The new calibration results are tested against the
reference calibration parameters listed under the REFERENCE column. The CURRENT
column shows the current operating parameters that are being used by the instrument when
calculating the analysis measurement result. Calibrations are similar to analysis cycles and are
on the method set in METHOD PARAMETERS menu.
Validation Parameters Menu
Allows configuration of parameters for validation, choice of validation source, and repetitions
required.
Analysis Parameters Menu
Allows configuration of parameters for instrument mode, measurement frequency,
measurement units, analysis label, cell wash and reagent flow rates, stability criteria and post
wash events. Separate post wash events can be programmed for analysis, calibration, and
validation cycles. All post wash cycles can be turned OFF when set to the "NOT ASSIGNED".
Cleaner Parameters Menu
Allows configuration of the parameters for manual or periodic cleaning of the reaction cell.
2230XP Silica Analyzer

Advertisement

Table of Contents
loading

Table of Contents