2230XP Silica Analyzer
Messages
No records to display
No records to print
Only one record to display
Problem with memory
allocation. Insufficient
memory in the system
or system is busy.
Measurement/Calibration/Validation Messages
Messages
Reagent1 empty
Reagent2 empty
Reagent3 empty
Reagent1, Reagent2 are
empty
Standard empty
Low standard empty
Hi standard empty
Two standards are empty
Sample empty
Testing optics, please check
for optics failure
Cause
User is trying to display records, and
no records are available in the
selected data range.
User is trying to print records, and no
records are available in the selected
data range.
User is trying to view graph when only
one analysis record is available in the
database.
Do not try to repeatedly view the
graph.
Cause
Performing Analysis/ Calibration/
Validation with two or three Reagents
and Reagent1 is exhausted.
Performing Analysis/ Calibration/
Validation with two or three Reagents
and Reagent2 is exhausted.
Performing Analysis/ Calibration/
Validation with two or three Reagents
and Reagent3 is exhausted.
Performing Analysis/ Calibration/
Validation with two or three Reagents
and Reagent1 & 2 are exhausted.
Performing single point or baseline
calibration and the Standard is
exhausted.
Performing two-point calibration
and the Low Standard is exhausted.
Performing two-point calibration
and the Hi Standard is exhausted.
Performing two-point calibration
and both standards are exhausted.
Performing analysis with single
sample and sample chamber is empty.
No light output from optics.
Maintenance Schedule
Section 9
Solution
None.
None.
None.
Try again after some duration or
after performing one more
measurement cycle.
Solution
Fill the empty Reagent1 bottle
with Reagent1.
Fill the empty Reagent2 bottle
with Reagent2.
Fill the empty reagent3 bottle
with Reagent3.
Fill the empty Reagent1 & 2
bottles with Reagent1 & 2.
Fill the empty standard bottle with
standard.
Fill the empty Low-Standard
bottle with Low-Standard.
Fill the empty Hi-Standard bottle
with Hi-Standard.
Fill the empty standard bottles
with their respective standards.
Check for Sample flow.
Check for pipe blockages
to sample chamber.
Run a Clean Cycle and try again.
Inspect cell and clean, check for
light detection under LED &
DETECTOR CONTROL in
Diagnostics menu.
Thermo Scientific Orion
73