Figure 7: Electrical wiring ports
Table 1: AC power terminal wire
connections
Figure 8: AC power connections
2230XP Silica Analyzer
BEFORE MAKING ANY CONNECTIONS:
Make sure that the wires are not powered before connecting to the instrument.
Disconnect any power previously connected to the relays.
Connect the Phase Hot or L (line), Earth and Neutral to the terminal block. Note the location
of the fuse.
Insert the wire into the terminal connector and make sure to tighten the screw.
Route signal connections to the back hub, and power connections separately to the front hub.
All connections must conform to local safety standards.
Analyzer Terminal Destination
HOT or L ( Line)
Protective Earth Ground (PG)
Neutral (n)
Installation Environmental Considerations
North America Wire Color
black
green
white
Section 3
European Wire Color
brown
green with yellow tracer
blue
AC POWER
CONNECTIONS