13. Insert the new filter and O-ring into the gas entrance block.
14. Replace the two screws that secure the gas inlet entrance assembly.
15. Lower the side panels of the pycnometer.
16. Replace the rear panel of the pycnometer and secure with the screws
Cleaning the Pycnometer
The exterior casing of the pycnometer may be cleaned using a clean cloth,
dampened with isopropyl alcohol (IPA), a mild detergent solution, or a 3%
hydrogen peroxide solution.
Do not immerse the pycnometer or the power cord in any liquids. Doing
so could result in electrical shock to personnel or damage to the unit.
Do not allow liquid to penetrate the casing of the pycnometer. Doing so
could result in damage to the unit.
Recovering from a Power Failure
The pycnometer has a battery back-up feature that saves entered data in case
of a power failure. Set up parameters and any other data entered will still be
present when power is restored. If an automatic operation was in progress
when the power failure occurred, it will be canceled when the pycnometer re-
starts. Any data collected during the automatic operation will still be present,
but the operation should be started again in order to produce complete re-
sults.