Queued Messages - Micromeritics AccuPyc 1330 Operator's Manual

Pycnometer
Hide thumbs Also See for AccuPyc 1330:
Table of Contents

Advertisement

AccuPyc 1330
Error and Status Messages
The following error messages may appear in the display area of the pycnome-
ter. Some of the messages contain the statement: "NN/ZZ runs completed."
When this appears, NN will be replaced with the number of runs completed
when the error occurred and ZZ will be replaced with the number of runs re-
quested.
The messages are organized into two categories: Queued Messages and Mes-
sages Displayed During an Automatic Operation.

Queued Messages

Messages placed in a queue are displayed in chronological order when you
press
next to the Reload prompt.
ANLSERR: Eq failure
NN/ZZ runs completed
ANLSERR: Fill failure
NN/ZZ runs completed
September 1996
. When the queue contains error messages, an asterisk is shown
CHOICE
Cause:
The sample being analyzed failed to equilibrate in 1000
seconds.
Action:
Check the system carefully for leaks as described in
Checking the Cell and Expansion Chambers for Leaks
later in this chapter. Make sure the sample is properly
prepared before performing an analysis.
Cause:
Sample (foams, organics, etc.) absorbs helium slowly.
Action:
Set the equilibration rate to progressively higher values un-
til reasonable equilibration times (15 sec to 120 sec) are
achieved.
Cause:
There was insufficient pressure to allow filling within five
minutes during an analysis.
Action:
Increase the regulator pressure or, if there is insufficient
helium in the tank, obtain a new tank of helium.