Reviewing And Editing Calibration Results And Parameters - Micromeritics AccuPyc 1330 Operator's Manual

Pycnometer
Hide thumbs Also See for AccuPyc 1330:
Table of Contents

Advertisement

AccuPyc 1330

Reviewing and Editing Calibration Results and Parameters

Press
September 1996
to review or edit calibration results.
5
Display
Cal std volume:
(volume)
Which chamber?
Cell
Cell [ N] = (volume)
Dev [ N] = (deviation)
or
Exp [ N] = (volume)
Dev [ N] = (deviation)
Reviewing and Editing Data
Entry
The volume of the calibration standard is
displayed.
Select either cell chamber or expansion
chamber, then press
ENTER
Either cell or expansion volume is
displayed depending on the selection made
at the previous prompt. [N] is replaced with
the number of the run. The deviation from
the mean is also displayed.
You may choose to exclude the displayed
cell or expansion volume from the
calculated average by pressing
When a volume is excluded from the
average, an asterisk appears next to it.
Likewise, a cell or expansion volume that
was previously excluded from the calculated
average may be included by pressing
.
When you press
CHOICE
calculated and displayed.
Press
to display values for the next
ENTER
run.
Press
when you wish to exit review
SAVE
and edit mode and return to display mode.
When you press
SAVE
are automatically recalculated and all data
reduction messages are added back into the
queue.
.
.
CHOICE
CHOICE
, a new deviation is
, all collected data
4-9