AccuPyc 1330
DTA_WRN: Cal std
10% of full-scale
DTA_WRN: Volume
10% of full-scale
HW_ERR: A/D I/O
failure
HW_ERR; BB RAM has
been initialized
HW_ERR: Valve I/O
failure
September 1996
Cause:
The system tried to generate a calibration report, but the
calibration standard used was less than 10% of the nomi-
nal cell chamber volume. Resulting data may be inaccu-
rate.
Action:
Use a calibration standard of sufficient size (calibration
standards are available from Micromeritics). The calibra-
tion standard should occupy at least 10% of the nominal
cell chamber volume and the more nearly filled the cell
is, the better the calibration.
Cause:
The amount of sample placed in the sample cup was so
small that the percentage of precision may be inferior.
Action:
Rerun the sample using enough material to occupy the
maximum amount of the nominal cell chamber volume.
Cause:
An error in the pressure measurement electronics has
occurred.
Action:
Call a Micromeritics Service Representative.
Cause:
Battery-backed RAM has failed.
Action:
Call a Micromeritics Service Representative.
Cause:
An error in the valve control electronics has occurred.
Action:
Call a Micromeritics Service Representative.
Error and Status Messages
5-9