Error and Status Messages
Messages Displayed During an Automatic Operation
When an automatic operation is in progress, status messages are continually
displayed.
Automatic operation
has been canceled
Print in progress
[ESCAPE] to cancel
Printer port
not responding
Sending line (line number)
[ESCAPE] to cancel
Transmission port
not responding
5-12
Cause:
The automatic operation has been canceled by the user.
Action:
Wait for the termination process to complete or end the
termination process by pressing
more times.
Cause:
Status message displayed when a report is being printed.
Action:
None; this is a status message only.
Cause:
The printer took longer than 30 seconds to acknowledge
receipt of data from the pycnometer.
Action:
Check to make sure the printer is properly connected to
the pycnometer, is turned on, and is on line.
Cause:
Status message displayed when data are being transmitted.
Action:
None; this is a status message only.
Cause:
The receiving device took longer than five seconds to ac-
knowledge receipt of data from the pycnometer.
Action:
Make sure the receiving device is properly connected to
the pycnometer RS-232 port and is turned on. Verify that
the serial I/O parameters controlling the receiving device
correspond with the data transmission parameters in the
set up mode.
AccuPyc 1330
two
CLEAR
March 2001
Need help?
Do you have a question about the AccuPyc 1330 and is the answer not in the manual?