Error and Status Messages
CAL2ERR: Overrange
NN/ZZ runs completed
CAL2ERR: Underrange
NN/ZZ runs completed
DATA_ERR:
No data to compute
5-8
Cause:
A pressure overrange occurred during the second pass of
calibration because the regulator pressure is set too high.
Action:
Adjust the pressure so that it is equal to the higher of the
two fill pressures specified in the analysis parameters plus
2.0 psig.
Cause:
A pressure overrange occurred during the second pass of
calibration because an error occurred in the pressure meas-
urement electronics.
Action:
Call a Micromeritics Service Representative.
Cause:
A pressure underrange occurred during the second pass of
calibration because an error occurred in the pressure meas-
urement electronics.
Action:
Call a Micromeritics Service Representative.
Cause:
An automatic operation was canceled before all necessary
data could be collected.
Action:
Restart the automatic operation.
Cause:
All the runs have been excluded in review mode.
Action:
Return to review mode and include some runs.