Analysis Or Calibration Parameters - Micromeritics AccuPyc 1330 Operator's Manual

Pycnometer
Hide thumbs Also See for AccuPyc 1330:
Table of Contents

Advertisement

AccuPyc 1330

Analysis or Calibration Parameters

Press
September 1996
2
to display or edit analysis or calibration parameters.
Display
Set-up type?
Analysis parameters
Number of purges:
3
Purge fill pressure:
19.500 psig
Number of runs:
3
Run fill pressure:
19.500 psig
Editing the Setup Parameters
Entry
Press
until Analysis parameters is
CHOICE
displayed. Then press
Enter the number of purges to be performed
and press
.
ENTER
Range:
0 to 99
Purging cleans the cell and expansion
chambers before an analysis begins. The
greater the number of purges, the cleaner
the sample will be when analyzed.
Enter the fill pressure and press
Range:
0 to 19.850 psig
0 to 136.86 kPag
Generally, the greater the fill pressure, the
easier it is to measure the volume precisely.
However, a lower pressure may be desirable
for some samples.
Enter the number of runs to be performed
and press
.
ENTER
Range:
1 to 99
Enter the fill pressure and press
Range:
0 to 19.850 psig
0 to 136.86 kPag
.
ENTER
.
ENTER
.
ENTER
4-19