Error and Status Messages
MAN_ERR:
Pressure overrange
PRT_ERR: Timeout
failed to respond
SYS_ERR: Power Fail
NN/ZZ Runs Completed
TRN_ERR: Timeout
failed to respond
USR_ERR: Cal std
10% of full-scale
5-10
Cause:
The fill valve was left open until the maximum system
pressure was exceeded.
Action:
Close the fill valve and open the vent and expansion
valves. Allow the pressure to stabilize.
Cause:
The printer took longer than 30 seconds to acknowledge
receipt of data from the pycnometer.
Action:
Check to make sure the printer is properly connected to
the pycnometer, is turned on, and is on line.
Cause:
A power failure occurred and when power resumed, the
automatic operation was canceled.
Action:
Restart the automatic operation if desired.
Cause:
The receiving device took longer than 10 seconds to ac-
knowledge receipt of data from the pycnometer.
Action:
Make sure the receiving device is properly connected to
the pycnometer RS-232 port and is turned on. Verify that
the serial I/O parameters controlling the receiving device
correspond with the data transmission parameters in the
set up mode.
Cause:
You tried to enter a calibration standard volume that is
less than 10% of the nominal full-scale volume.
Action:
Enter a volume that represents at least 10% of the nomi-
nal cell chamber volume.
AccuPyc 1330
March 2001