Micromeritics AccuPyc 1330 Operator's Manual page 72

Pycnometer
Hide thumbs Also See for AccuPyc 1330:
Table of Contents

Advertisement

AccuPyc 1330
USR_ERR: No data to review
USR_ERR: Number of runs must be = 5
USR_ERR: Out of range
USR_ERR: Pressure overrange
ZEROERR: Eq failure
ZEROERR: Overrange
ZEROERR: Underrange
March 2000
Cause:
You tried to review data for an automatic operation when
there were no data to review.
Action:
Abandon request.
Cause:
You tried to enable run precision without increasing the
number of runs to at least five.
Action:
Increase the number of runs to five or abandon request.
Cause:
You tried to enter a value that is out of the valid range.
Action:
Enter a value in the specified range (refer to Chapter 4).
Cause:
A pressure overrange occurred but was left uncorrected.
Action:
Return to manual mode and vent the system.
Cause:
The system failed to equilibrate within 1000 seconds.
Action:
Check the pycnometer for leaks. Make sure the pycnome-
ter has been placed in a draft-free environment.
Cause:
A pressure overrange occurred while zeroing.
Action:
Check for a fill valve leak. If there is none, contact a
Micromeritics Service Representative.
Cause:
An error occurred in the pressure measurement electronics
while zeroing, causing a pressure underrange to occur.
Action:
Call a Micromeritics Service Representative.
Error and Status Messages
5-11