7-4-4-3

Connected Probes

Connected Probes shows probes connected to the system.
Note:
The information on Connected Probes is available to all service class licenses.
The following table lists all the elements available on Connected Probes section:
Table 7-29
Element
Active Probe Tempera-
ture (Celsius)
Probe Name
Probe ID
Serial Number
Status
7-4-4-4

Gather Logs

Gather Logs provides a way to collect system logs and place the log files in the D:\Service directory for
retrieval by the online center. Log files are compressed into a zip file and the file path and name display.
Note:
The Gather Logs is available to all service class licenses.
To access the Gather Logs utility:
1.) Tap Utility on the top bar and select Gather Logs
2.) Select one of the following:
Figure 7-14 Connected Probes
Connected Probes
When available, temperature of the active probe.
Relevant for TEE probe.
Name of the probe connected to the system
Identifier of the probe connected to the system.
Serial number of the probe connected to the system. If the serial number of
the probe is not available, then N/A will be displayed.
Statuses of the probe connected to the system. Valid values are:
• Active
• Non Active
• 1 Day Logs
• 1 Week Logs
• All Logs