Speakers Replacement Procedure; Tools; Time Required; Preparations - GE Venue Fit Technical Publication

Table of Contents

Advertisement

P R E L I M I N A R Y
DIRECTION 5854713-100, REVISION 2
8-2-5

Speakers Replacement Procedure

8-2-5-1

Tools

Phillips screwdrivers of appropriate size
8-2-5-2

Time Required

5 min
8-2-5-3

Preparations

1.) Shut down the Venue Fit™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Fit™ system from cart as described in
and place it on a previously prepared surface with the display facing down.
8-2-5-4

Speakers Removal Procedure

1) Remove back cover.
2) Disconnect the speakers connector from the M-BEB.
3.) Remove the 2 grounding screws marked with red and green circles (the screw marked in red is
longer) and release the 2 supporting screws (marked with blue circles)
Rear Cover Removal Procedure
Figure 8-38 Speakers removal
Chapter 8 - Replacement Procedures