Impact shock detection (ISD) function
This function detects impact shocks to the mass sensor section, displays the impact level, and records
it. (Available with display unit software versions 1.010 or later)
By lowering the impact level at the time of loading, it is possible not only to alleviate variation in the
weighing value but also to reduce the risk of failure of the mass sensor section. Especially when
incorporating the balance in a production line, etc. and weighing by means such as an automated
system, impact greater than expected may be applied to the sensor. When designing automatic
systems and the like, it is recommended that you minimize the impact level as much as possible while
checking the shock indicator.
Impacts of impact level 3 or higher are stored on the balance with date and time.
For details, refer to
You can hide the impact level by turning off the Warning display button.
If the impact level is level 3 or higher, the record will be kept automatically even if the impact level is
hidden.
Caution
Impact on the weighing sensor is not only that applied to the weighing pan when loaded, but also may
be impact applied from the table on which the balance is installed. The impact detection function also
works for impact coming from the table.
The shock indicator has 5 levels from level 0 to level 4.
Impact level
No indicator
Level 0
Level 1
Level 2
Level 3
Level 4
"13.29 Impact shock detection
Shock
Buzzer
indicator
No beeps
One beep
Two beeps
history". Impact shock detection history".
Safe
Caution
Caution:
Alleviate impact shocks
Warning:
Do not apply any more impact shocks
Danger:
Sensor may be damaged