21. UFC Function
The UFC (Universal Flex Coms) function allows you to output contents of your choice when outputting
the weighing data. You can also output a character string when printing a barcode with a label printer or
the like.
In order to use the UFC function, it is necessary to set the output data format used to [UFC format]. For
the data format settings, refer to
LAN",
and
"13.16. Bluetooth".
UFC program commands
The output data format used can be stored in the balance by sending the program command from a
personal computer.
Text files can be imported from a USB flash drive to the internal memory.
For the import method, refer to
The
program
"Received_Program_Command.txt". The data in the internal memory is stored even when the power of
the balance is turned off. However, when the program command is received again, the file will be
overwritten.
Creating program commands
The maximum number of characters for a program command is 1024 characters.
Add the three characters "PF," to the beginning of the program command.
Program commands are combined using comma or space delimiters, which can be omitted to reduce
the number of characters. The comma after the PF command, however, cannot be omitted.
Program command list
Command
UFC command header
Added to the beginning of the program
PF,
command.
Manufacturer name
$MN
Model name
$TY
$SN
Serial number
ID number
$ID
$DT
Date
Time
$TM
$WT
Weighing data
$GR
Gross data (gross weight)
Net data (net weight)
$NT
$TR
Tare data (tare weight)
Counting data
$PC
$UW
Unit weight data
Comma
$CM
Space, ASCII 20h
$SP
$CR
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
$LF
Outputs a weighing value in the NU2 format
$NU
$HT
Outputs a tab
*
Supported with the touch panel software version 1.019 or later
"13.13. RS-232C interface", "13.14. USB interface", "13.15.Wired
"13.17. UFC
command
received
Function
*
format".
is
saved
in
the
S A M P L E - 1 2 3 4 - 5
2 0 2 1 / 1 0 / 0 1
1 2 : 3 4 : 5 6
+ 0 . 2 3 4 5 6 7
+ 1 . 2 3 4 5 6 7
+ 0 . 2 3 4 5 6 7
+ 1 . 2 3 4 5 6 7
,
C
R
L
F
0 . 2 3 4 5 6 7
*
TAB
199
internal
memory
with
Output example
A
&
B A - 6 T E
T 1 2 3 4 5 6 7
+ 1 2 3 4
+ 0 . 1 2
the
filename
D
g
g
g
g
P C
g

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte

Table of Contents