A&D BA-T Series Instruction Manual page 44

Analytical balance
Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Minimum weight setting for measurement
Display settings: HOME key
Application selection button
HOME key
Minimum weight setting button
Input by measuring repeatability button
1
Measurement method
2
Measurement tolerance
Name
Minimum weight
1
measurement
method
Minimum weight
2
measurement
tolerance
3
Measure button
Settings in the red box are default values (factory settings).
This is the screen for measuring the minimum weight.
Automatic (internal weight)
For this setting, the minimum weight is automatically measured with the internal weight in the
[Repeatability measurement mode] screen. When the measurement is complete, the [Minimum weight
measurement result] screen described in
1
Back button
Name
Back button
1
APP button
Weighing screen with the minimum weight alert function
Setting value
Automatic
(internal weight)
Select the minimum weight measurement method.
Manual (external
weight)
0.1
%.
Select the minimum weight measurement tolerance.
1.0
%.
Start the minimum weight measurement.
When the measurement is complete, the [Minimum weight
measurement result] screen described in
weight measurement
"5.5.4. Minimum weight measurement
Returns to the previous screen.
[Application] settings screen
Select [Minimum weight alert function]
[Minimum weight setting] screen
Description
result" is displayed.
Description
44
3
Measure button
"5.5.4. Minimum
result" is displayed.