A&D BA-T Series Instruction Manual page 147

Analytical balance
Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Command settings
Display settings: MENU key
1
Terminator
2
AK (acknowledge), error code
3
Command time out
4
Back button
Name
1
Terminator
AK (acknowledge),
2
error code
3
Command time out
4
Back button
Settings in the red box are default values (factory settings).
C
: Carriage return, ASCII 0Dh
R
When [AK (acknowledge), error code] is set to ON in the command settings, the balance always
responds to reception of all commands sent from a personal computer or a PLC. Checking the code that
is responded improves the reliability of the communication.
Balance response
When [AK (acknowledge), error code] is set to ON in the function table, the balance responds as
follows.
When a command requesting data is sent to the balance:
If the balance cannot send the requested data, it sends an error code (EC,Exx).
If the balance can output the requested data, it sends the requested data.
When a command controlling the balance is sent to the balance:
If the balance cannot execute the command, it sends an error code (EC, Exx).
If the balance can execute the command, it sends the AK code (acknowledgment, ASCII 06h).
Command
ON
Turns the display on.
P
Turns the display on/off. (Only when the display is on.)
Same as the RE-ZERO button
R, RZ
Same as the TARE button
T, TR
Zero
ZR
Sensitivity adjustment with the internal weight.
CAL
EXC
Sensitivity adjustment with an external weight.
System settings button
Communication button
Command settings button
Setting value
(setting range)
C
L
, C
R
F
R
OFF , ON
No limit,
Limit to one second
-
If the load is within ±2% of the capacity from the initial zero point, the zero
point is updated, the tare value is cleared and the display is set to zero. If the
load exceeds ± 2%, no processing is done.
Data output button
[Command settings] screen
Selects the terminator for the output data.
Select the response (received, processing or process
completed) to all commands sent from PC or PLC.
Sets the command timeout.
Returns to the previous screen.
Description
.
.
147
Description
AK : Acknowledgment, ASCII 06h
L
: Line feed, ASCII 0Ah
F

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte

Table of Contents