A&D BA-T Series Instruction Manual page 148

Analytical balance
Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

GLP output
Display settings: MENU key
1
GLP output
2
Clock to be used
3
Back button
Name
GLP output
1
Clock to be
2
used
Back button
3
Settings in the red box are default values (factory settings).
Main objectives
The GLP/GMP compliant data can be output to a personal computer or optional printer.
The GLP/GMP compliant report includes the balance manufacturer (A&D), model name, serial number,
ID number, date, time, and space for a signature. For a sensitivity adjustment or calibration test, the
result and the weight used are also included.
The balance can output the following GLP/GMP compliant reports via the connected external output.
Sensitivity adjustment report (Output for sensitivity adjustment using the internal weight, automatic
sensitivity adjustment, sensitivity adjustment using the external weight)
Calibration test report (Output for calibration test using an internal weight or external weight)
Breaks ("title block" and "end block") for easy management of a series of weighing data
System settings button
GLP output button
Setting value
(setting range)
OFF , ON
Balance internal clock ,
External device clock
-
[GLP output] screen
Change the setting to output GLP.
Sets the clock for GLP output.
Returns to the previous screen.
148
Communication button
Description

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte

Table of Contents