Sensitivity adjustment setting
Display settings: MENU key
[Sensitivity adjustment setting] screen
1
External weight value input
2
Enter the correct internal weight value button
1
External weight value input
Enter the correct internal weight value
2
button
Settings in the red box are default values (factory settings).
*1
Input range
BA-6TE/BA-6DTE
BA-225TE/BA-225DTE
BA-125DTE
Correcting the internal weight value
Display settings: MENU key
Enter the correct internal weight value button
1
Back button
Name
1
Back button
This function corrects the internal weight value based on your external weight. Execute the external
sensitivity adjustment in advance.
The balance automatically loads and unloads the internal weight and corrects the internal weight value.
Then the balance automatically executes the internal sensitivity adjustment. When the adjustment is
complete, the [Sensitivity adjustment setting] screen is displayed.
The corrected value is stored in nonvolatile memory even if the AC adapter is removed.
Sensitivity adjustment button
Name
: 0.9 g ~
: 9.9 g ~ 200 g ~ 200.1 g
: 9.9 g ~ 100 g ~ 100.1 g
Sensitivity adjustment button
Returns to the previous screen.
Setting value
(setting range)
Input the external weight value. This is
common to the "Adjustment weight value"
Input range
*1
input item in the [External sensitivity
adjustment] screen.
Displays the [Internal sensitivity
adjustment] screen.
5 g ~
5.1 g
[Internal sensitivity adjustment] screen
Description
100
Settings button
Description
Settings button