Step 5.
Enter the [Molecular formula] and [Sample name] to be
registered. The figure on the right shows an input
example on the [Sample registration] screen.
After inputting, press the Back button to display the
[Sample edit] screen.
Enter the fields then press the Back button
Step 6.
Register the first sample in the recipe. Select the desired
sample using the Selection button, and then enter
values in the [Target value] and [Tolerance] fields.
In the example, the sample name is Na2HPO4, the target
value is 10.000 mmol/L, and the tolerance is 10.000 %.
After inputting, press the right arrow button
bottom right of the screen to display the registration
screen for the second sample.
The right arrow button
Step 7.
Enter the contents of the remaining samples in the same way.
Then, press the Register button to register the recipe and to
display the [Recipe edit] screen.
Caution: If the Back button, ON:OFF key
key
, LOG-IN key
pressed to display another screen, the sample is not
registered in the recipe.
Step 8.
Now the recipe registration is complete.
To start measurement with the registered recipe,
press the Measure button on the [Recipe edit]
screen.
To register another recipe, press the Back button on
the [Recipe edit] screen to display the [Recipe list]
screen, and then register the recipe on the [Recipe
list] screen shown in step 2.
displays the registration screen for the next sample.
, or MENU key
Press the Register button
at the
, HOME
is
70