BA-6TE / BA-6DTE
BA-225TE / BA-225DTE / BA-125DTE
Minimum weight alert function
If you configure this display, the HOME screen changes to the weighing screen with the minimum
weight alert function.
Display settings: HOME key
1
Minimum weight alert function
2
Minimum weight setting button
Name
Minimum weight alert
1
function
Minimum weight setting
2
button
Minimum weight alert
3
display
This is the screen for the minimum weight alert function. This function can be used only with the unit
"mg".
Minimum weight alert function usage
Minimum weight is the minimum sample weight required to perform correct quantitative analysis taking
the measurement error of the balance used into account. If the sample amount is too small, the
proportion of measurement error in the measured value increases, and the reliability of the analysis
result thus may drop.
The minimum weight alert function makes it possible to judge immediately whether the sample amount
meets the set minimum weight.
Model
APP button
Application selection button
HOME key
Weighing screen with the minimum weight alert function
Displayed when the minimum weight alert function is enabled.
Displays the [Minimum weight setting] screen.
Also displays the currently stored minimum weight.
Displayed when the sample weight is less than the set minimum weight.
100% reference mass
0.0010 g -
0.0099 g
0.0100 g -
0.0999 g
0.1000 g -
0.0100 g -
0.0999 g
0.1000 g -
0.9999 g
1.0000 g -
[Application] settings screen
Select [Minimum weight alert function].
Description
42
Decimal separator
position
1 %
0.1 %
0.01 %
1 %
0.1 %
0.01 %
3
Minimum weight alert display