6. Sensitivity Adjustment
The sensitivity adjustment is performed. When the adjustment is complete, the [Sensitivity adjustment
result] screen is automatically displayed.
Be sure to warm up the balance with nothing on the weighing pan for at least an hour, or at least
four hours for BA-6TE/BA-6DTE, with the AC adapter connected to the power supply.
When the internal weight is used, this is automatically performed. When an external weight is used,
follow the instruction.
Do not apply vibration and the like to the balance during sensitivity adjustment.
1
2
3
4
Name
Weight to be used
1
Instruction display
2
External weight
3
value input
Back button
4
Instruction display
5
Skip button
6
Execute button
7
Result display
8
PRINT button
9
Settings in the red box are default values (factory settings).
*1
Input range
BA-6TE/BA-6DTE
BA-225TE/BA-225DTE
BA-125DTE
Setting value
(setting range)
Internal weight ,
Select the weight to be used.
External weight
Displays the instruction for external sensitivity adjustment.
Input the external weight value.
Input range
*1
This is common to the external weight value input field in the
[Sensitivity adjustment setting] screen.
-
Cancels and returns to the previous screen.
Displays the progress of the internal sensitivity adjustment.
Skips this check item and proceeds to the next item.
Executes the sensitivity adjustment.
Displays the sensitivity adjustment result.
Outputs the results to the device connected to the balance.
The results are automatically output if [GLP output] in [Data
output] is set to ON.
: 0.9 g ~
5 g ~
: 9.9 g ~ 200 g ~ 200.1 g
: 9.9 g ~ 100 g ~ 100.1 g
5
6
7
8
9
Description
5.1 g
111