A&D BA-T Series Instruction Manual page 46

Analytical balance
Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Minimum weight measurement result
Display settings: HOME key
Application selection button
HOME key
Minimum weight setting button
Input by measuring repeatability button
Measure button
The [Minimum weight measurement result] screen is displayed when the
for measurement
1
Measurement method
2
Measurement tolerance
3
Repeatability (standard deviation)
4
Minimum weight
Name
1
Measurement method
2
Measurement tolerance Displays the minimum weight measurement tolerance.
Repeatability (standard
3
deviation)
4
Minimum weight
5
PRINT button
6
Register button
*
If the repeatability (standard deviation) is 0.41d or less, the minimum weight is 2000 times the 0.41d in
compliance with USP.
"d" is a unit of readability.
APP button
Weighing screen with the minimum weight alert function
[Minimum weight measurement result] screen.
is complete.
Displays the method used for the minimum weight measurement.
Displays the standard deviation for this measurement.
Displays the minimum weight.
Outputs the minimum weight measurement result to the device connected to
the balance.
Registers the minimum weight.
[Application] settings screen
Select [Minimum weight alert function]
[Minimum weight setting] screen
Description
46
5.5.3.
Minimum weight setting
5
PRINT button
6
Register button