Sensitivity adjustment while the keys are locked
The following commands are enabled on the sensitivity adjustment screen while the keys are locked.
Command string
NEXT
BACK
Procedure for external sensitivity adjustment while the keys are locked
Weighing example for BA-6DTE
Step 1.
Make sure that nothing is on the weighing pan,
and then send the [NEXT] command. The balance
measures the zero point.
Do not apply vibration and the like to the balance.
You can input the external weight value before input.
Step 2.
Place the weight on the weighing pan, and then
send the [NEXT] command.
Measure the weight.
Do not apply vibration and the like to the
balance.
You can input the external weight value before input.
Step 3.
The [Sensitivity adjustment result] screen for the
external sensitivity adjustment is automatically
displayed. Please remove the weight.
Step 4.
The result screen is displayed.
The result is automatically output if [GLP output] in
[Data output] is set to ON.
Step 5.
Send the [BACK] command to return to the weighing screen.
Place the weight again to confirm that the sensitivity of the balance is adjusted properly.
If it is not within the range, start over from the first step of this procedure in the appropriate
ambient conditions.
Confirms the current input and proceeds to the next instruction.
Operates the same as the Enter button
Displays the [HOME] screen.
Operates the same as the Back button.
Description
.
Weighing pan
[NEXT ] command
External weight
[NEXT ] command
GLP output
202
Result display